Skip to content

Specifications for XE-8 | Q-SUN Xenon Arc Large Capacity Rotating Rack Tester

Dimensions

  XE-8-HBSE
External 
WxDxH (mm) 1730 × 2040 × 1040
Specimen Area
h x dia = Rack Area 77.5 x 47.5 cm = 11,560 cm2
Specimen Capacity
QTY @ size 164 @ 102 x 51
Maximum Specimen Weight
Distributed Evenly (kg) 68
Weight
Net Weight (kg) 518

 

Technical Specification

  XE-3-HBSE

Black Panel Temp (°C)    

Light    

Dark 

BP                     IBP

39-100                  39-110

25-45                    25-45

Chamber Air Temp (°C)    

Light    

Dark 

CAT

35-65

25-45

Relative Humidity 20-95%
Inlet Water Pressure and Purity

2.1-6.2 bar (30-90 psi); > 5 MΩ∙cm; < 0.2 μS /cm; < 0.1 ppm TDS; < 0.1 ppm colloidal silica

Water Consumed with Spray On 1.0 L/min (front)
1.5 L/min (front & back)
Water Consumed with Humidifier On

0.4 L/min at 100% duty cycle

ISO 4892-2 Cycle 2: (30% duty cycle) = 173 L/day

Electrical Requirements 3Ph/50A/400V

 

Features

  XE-8-HBSE
Chamber Type Rotating Rack
Specimen Capacity (41 holders × 4 specimens per holder) 51 × 102 mm 164
Specimen Orientation (measured from horizontal) 90°
Dual Touch-Screen Displays in 17 Languages ✔
Full Spectrum, Ozone-Free LIGHT/YEAR® Xenon Arc Lamps - 4200 W (replace ~once/year) 4
Longer Lamp Lifetimes and/or Higher Irradiance  ✔
Long-Life Optical Filters ✔
Type I and Type II Daylight Filter Options ✔
SOLAR EYE Irradiance Control (340 nm, 420 nm or TUV)  ✔
Dual Uninsulated and Insulated Black Panels for precise control of temperature uniformity ✔
Chamber Air Temperature Control ✔
Relative Humidity Control ✔
Water Spray (Front and Back) ✔
AUTOCAL UC80 Ride-Around Universal Calibrator System ✔
Chiller Optional
USB Port for Software Updates and Tester Data Acquisition ✔